68 / 106 HEYER Pasithec, Operator’s manual Rev. 0.2 Draft - 12/09
7.3.3 Regular Maintenance
WARNING: To avoid patient injury, do not perform any tests or repairs when the
equipment is in use.
Perform the following check every 30 days to ensure that component worn by use and daily cleaning
are replaced in a timely manner.
Test by eyes
Separate the bellows assembly and anesthesia machine.
Disassemble the bellows assembly
WARNING: Never separate the diaphragm and the valve seat in a pressure-relief
valve.
Check each component carefully to check for cracks, distortion, dissolution, inflation and other
physical changes. Replace them if necessary.
Assemble the bellows assembly, and then perform the leak test.