5.4 Installing and Testing Vaporizer
5.4.1 Installation
WARNING: Do not remove the vaporizer from the bypass valve with its locking lever
locked.
Do not use more than one vaporizer simultaneously with this system.
Install vaporizers by following these steps:
1. The vaporizer must be disassembled and reinstalled if its top is not horizontal.
2. Set the locking lever of the vaporizer so that it is locked.
3. Lift the vaporizer straight up to separate it from the bypass valve, but do not pull the vaporizer
forward. Be careful not to rotate it on the bypass valve.
4. As the vaporizer separates from the bypass valve, reinstall the vaporizer and then follow step 1 to
step 3. Do not use this system if it is not positioned horizontally on the bypass valve.
5. Attempt to open two vaporizers at the same time. If more than one vaporizer can be opened at
the same time, disassemble and reinstall them. Then perform step 1 to step 5.
5.4.2 Testing Vaporizer
CAUTION: Refer to relevant instructions for use regarding the performance testing of
the vaporizer.