Rev. 0.2 Draft – 12/09 HEYER Pasithec, Operator’s manual 45 / 106
5 Pre-use Check List
5.1 Pre-use Check List Procedure
Test interval: Pre-operative Checkout should be done each day prior to use by the first patient; prior
to the use of each subsequent patient, and after repair or maintenance.
The test schedule appears in the table below:
Prior to use by the first patient each day Prior to use by each patient
System check
Power failure alarm test
Gas pipeline and gas cylinder test
Flow control test
Vaporizer installation and test
Alarm test
Breathing system test
Ventilator test
Breathing System test
Ventilator test
WARNING: Do not use this system until the operation and maintenance manuals have
been read and understood.
Whole system connection
All warnings and cautions
User guide for each system module
Testing method for each system module
Before using this system:
Complete all tests of this section
Test all system modules
If a test fails, do not use this system. Please contact a service representative.
5.1.1 Check System
WARNING: Ensure the breathing circuit is connected correctly and in good condition.
Make sure:
1
Equipment is in good condition.
2
All the components are correctly connected.
3
Breathing circuit is correctly connected and in good condition and the breathing system contains
sufficient absorbent.
4
Vaporizer is in lock position and is filled with sufficient anesthetic.
5
The connection and pressure of pipeline gas supply system are correct.
WARNING: Do not leave the cylinder valves open during pipeline gas supply period.
The cylinder gas supply could be emptied and lead to insufficient supply in
case of pipeline malfunction.
6
The required emergency device is ready and in good condition.
7
The device for airway maintenance and organ cannula are ready and in good condition.
8
The applicable anesthetic and emergency medicine are ready.
9
Tighten and lock the truckles to ensure they are free of motion.
10
Connect the power cord to the AC power outlet. The power indicator light will glow when power is
connected. If the indicator does not glow, power is not being supplied to the machine. Plug the
power cord into a different socket, close the breaker, or replace the power cord.