H E Y E R P a s i t h e c
74 / 106 HEYER Pasithec, Operator’s manual Rev. 0.2 Draft - 12/09
8.3.2.2 Calibrate 100% O
2
Sensor
This item must be performed by an authorized service engineer.
8.3.3 Calibrate CO
2
Sensor
Step 1
Turn the knob to select “CO
2
sensor Calibration” in the “Calibrate” submenu.
Step 2
Press the knob and a new window appears on the right side.
Perform the operation according to the prompts on the screen.
Disconnect the Gas sensor from the Y-piece, and place it in the air for no less than 15 minutes.
Remove the airway adapter (or a new one) until the light turns green.
Wait 1 minute till the State-Area turns green (OK) and select it.