EasyManua.ls Logo

Heyer LUVAR A - Maintenance; Disposal

Default Icon
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
Page 75
heyer_LUVAR_UserManual_EN-3_CRev-1
Interference re-
sistance tests
IEC 60601 test
level
Compliance level Electromagnetic envi-
ronment - guidelines
Magnetic fields
at supply fre-
quency (50/60
Hz) as per IEC
61000-4-8
30 A/m 30 A/m Magnetic fields at
the mains frequency
should correspond
to the typical values
found in business and
hospital settings.
Voltage dips,
short-term
interruptions,
and voltage
fluctuations
of the power
supply lines
according to
IEC61000-4-11
Dip levels /
duration:
30%/500ms
100%/20ms
100%/10ms
at 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
100%/5000ms
Dip levels /duration:
30%/500ms
100%/20ms
100%/10ms
at 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
100%/5000ms
The quality of the
supply voltage should
correspond to that of
a typical business or
hospital setting.
If the device needs to
continue operating
uninterrupted in the
event of a power fail-
ure, the device should
be powered by an
uninterruptible power
supply or a battery.
Maintenance
The expected service life of the device is 5 years. Some components and accesso-
ries have a shorter service life: 2.5 years for the water tank and 12 months for the
breathing tube and mask.
The device is maintenance-free within this period if used as intended in accordance
with the instructions for use and if the cleaning and care instructions are observed.
For use beyond this, we recommend an inspection by an authorized service compa-
ny.
Disposal
Dispose of the device and packaging at the end of its service life in accordance
with local laws and regulations. Proper disposal conserves natural resources and
prevents harmful substances from entering the environment.

Table of Contents

Related product manuals