4.7 Speaker
Chapter 4: Removal, Replacement, and Adjustment Procedures
15. Remove the speaker (J) from the siderail (D).
16. Disconnect the wires from the speaker (J).
Replacement
1. Make sure the gasket (K) is installed in the siderail.
2. Do the removal procedure in reverse order.
3. To make sure the VersaCare™ Bed operates correctly, do the “Function
Checks” on page 2-6.
4
VersaCare™ Bed Service Manual (MAN333 REV 2) Page 4 - 23