Function Checks
Chapter 2: Troubleshooting Procedures
Function Checks
1. The "Initial Actions" have been performed.
Yes No
↓ → Go to “Initial Actions” on page 2-2.
2. Operate the trendelenburg function.
This function works.
Yes No
↓ → Go to RAP 2.1 on page 2-13 or go to RAP 2.16 on page 2-28.
3. Lock out the head, knee, hilow, and Master controls.
The corresponding INDICATOR illuminates.
Yes No
↓ → Go to RAP 2.1 on page 2-13 or go to RAP 2.16 on page 2-28.
4. Unlock the head, knee, hilow, and Master controls.
The corresponding INDICATOR goes out.
Yes No
↓ → Go to RAP 2.1 on page 2-13.
5. Press the Bed Up control of the hilow function.
The bed rises to the high position without stopping.
Yes No
↓ → Go to RAP 2.1 on page 2-13 or go to RAP 2.16 on page 2-28.
6. Press the Bed Down control of the hilow function.
The bed goes down to the low position without stopping. In addition, the
Bed Not Down INDICATOR goes off when the bed reaches the maximum
low position.
Yes No
↓ → Go to RAP 2.1 on page 2-13 or go to RAP 2.16 on page 2-28.
7. Press the Knee Up control of the knee section function.
The knee section rises to the high position without stopping or the audible
alarm sounding during the movement.
Yes No
↓ → Go to RAP 2.1 on page 2-13 or go to RAP 2.16 on page 2-28.
8. Press the Knee Down control of the knee section function.
Page 2 - 6 VersaCare™ Bed Service Manual (MAN333 REV 2)