4.15 Setting and Activating the Scaling
────────────────────────────────────────────────────
NOTE
4. Set the compensation coefficient b as the same way.
The setting can be made in the range : -200.00M to +200.00M
:
step the units up
:
step the units down
p--n --μ --m--(none)--k--M
<= Units down Units up =>
5. When the compensation coefficient setting has been established, press the
key on the Compensation coefficient setting screen to return to the
Scaling screen.
The scaled measurement value has no effect on other measurement values.
Therefore, even if the same parameter is set for the first and third
parameters, the scaling is carried out with the respective compensation
coefficients.
The calculation result of scaling for
θ
is as follows.
When exceeding 180°, "180.00°" is displayed.
When below -180°, "-180.00°" is displayed.
The setting can be made in the range : -999.99% to +999.99%