4.5 Setting the Voltage/Current Limit
────────────────────────────────────────────────────
4.5.1 Control Screen Sequence
Initial screen
Menu screen
Limit setting screen
Initial screen
4.5 Setting the Voltage/Current Limit
Depending on the test signal level, in some cases it is possible to damage the
sample which is being tested by applying to it a voltage or a current greater
than its rated value. Using the voltage/current limit function, it is possible
to set a limit value which the voltage applied to the sample under test, or
alternatively the current flowing through it, should not exceed; and thereafter
the 3522-50 unit will limit the voltage, or current, so as keep it below this
specified limit value.
When open circuit voltage (V) or constant voltage (CV) is set:
A current limit can be set.
When constant current (CC) is set :
A voltage limit can be set.
Press the key.
Press the key.
Input a limit value.
(The 3522-50 will automatically switch between
voltage limit and current limit according to the test
signal level setting.)
When the setting has been established, press the
k
ey to return to the Initial screen.