15
────────────────────────────────────────────────────
3.3 Basic Measurement
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
NOTE
3.3.1 Basic Flow up to Testing of the Sample
Step Reference sections
Select the test parameters to be
displayed
3.3.2, "Setting the Test Parameters to be Displayed (Cs, D)"
Set the test frequency
3.3.3, "Setting the Test Frequency"
Set the voltage level
3.3.4, "Setting the Constant Voltage Level"
Open circuit compensation
3.3.5, "Setting Open Circuit Compensation"
Short circuit compensation
3.3.6, "Setting Short Circuit Compensation"
Start testing
3.3 Basic Measurement
In order to explain the basic operation of the 3522-50 unit, as an example,
the procedure will be shown for establishing the following settings:
Example
Sample to be tested
Capacitor 85 μF
Test conditions
Test frequency 120 Hz
Constant voltage level 0.4 V
Open circuit compensation setup ALL compensation
Short circuit compensation setup ALL compensation
Parameters to be displayed
Capacitance Cs, Loss coefficient D
The trigger is internal. The 3522-50 unit, when it is dispatched from the
factory, is in the internal trigger state.
The basic flow up to starting testing of the sample is as shown in the
following chart in correspondence to the applicable reference sections: