80
────────────────────────────────────────────────────
4.13 Setting the Testing Speed
────────────────────────────────────────────────────
Test speed setting Maximum display digits
FAST 4
NORM, SLOW, SLOW2 5
4.13.1 Control Screen Sequence
Initial screen
Initial screen
Menu screen
Test speed setting screen
4.13 Setting the Testing Speed
The testing speed can be set. The slower the testing speed is, the more
accurate are the results.
The testing speed can be set to one of three values:
FAST
low accuracy testing is performed at high speed
NORM
the speed used for normal testing
SLOW
high accuracy testing is performed slowly
SLOW2
high accuracy testing is performed slowly
The number of displaying digit of the test values for parameter excluding D,
Q, and
θ
varies depending on the setting of test speed.
Press the key.
Press the key.
Select the speed for testing.
When the setting has been established, press the
key to return to the Initial screen.
The setting has been made.