135
────────────────────────────────────────────────────
5.5 The EXT I/O Connector
────────────────────────────────────────────────────
Signal
Page
PNL2
―――――
PNL1
―――――
PNL0
―――――
Invalid HIGH HIGH HIGH
1 HIGH HIGH LOW
2 HIGH LOW HIGH
3 HIGH LOW LOW
4 LOW HIGH HIGH
5 LOW HIGH LOW
6 LOW LOW HIGH
Invalid LOW LOW LOW
Page
Signal
1 2 3 4 5 6
LD1
――――
1 6 11 16 21 26
LD2
――――
2 7 12 17 22 27
LD3
――――
3 8 13 18 23 28
LD4
――――
4 9 14 19 24 29
LD5
――――
5 10 15 20 25 30
(8) TRIG
___________
When the external trigger is enabled for the 3522-50, measurement is begun
when this signal is input.
(9) PNL0
---------------
to PNL2
---------------
(Panel page signals)
These lines inputs the page which have been saved by the panel save
function.
(10) LD1
------------
to LD5
------------
(Panel load signals)
Select the panel numbers stored using the panel save along with PNL0
--------------
to
PNL2
--------------
.
This signal is used to select panel numbers stored using the panel save
function. When a panel number has been selected using this signal at the
input of a trigger signal, the 3522-50 loads the measurement conditions for
the panel number and then begins measurement. The table below shows the
selectable panel numbers.