22
────────────────────────────────────────────────────
3.3 Basic Measurement
────────────────────────────────────────────────────
Level Setting Screen
Level Setting Screen
Level Setting Screen
NOTE
Initial Screen
3. Because you for example want to set the value
for the constant voltage (CV), press the
key. Then check that the key changes to
black figures on a white ground (i.e., in non-
reversed video), which confirms that constant
voltage has been properly selected.
4. Using the
and
keys, set the displayed
constant voltage value to 0.400.
The corresponding digit is incremented.
(If you hold this key down continuously,
the corresponding digit increments
continuously.)
The corresponding digit is decremented.
(If you hold this key down continuously,
the corresponding digit decrements
continuously.)
5. When the setting has been established as above,
press the key.
With this constant voltage (CV) setting, the
testing will be performed with the voltage
between the terminals of the sample being
tested being kept constant at 0.400 V.
The accuracy of this constant voltage is
10% 10 mV.
For details, refer to Section 4.4, "Setting the Test
Signal Level"