96
)XQFWLRQDO6SHFL¿FDWLRQV
(3) Measurement-value Retaining function
Functionality The function retains (suspends update of) displayed values.
How to set Press the HOLD key.*
1
Input signals to the EXT.HOLD terminal.*
2
Measurement values stabilizes (when the auto-hold
function is on).
How to cancel Perform either *1 or *2 above. (toggle action)
(4) Measurement-value Auto-hold function
Functionality The function retains resistance measurement values
automatically once they get stabilized.
Default OFF
How to set Use the A.HOLD/MEMO key to turn it on and off.
(5) Comparator function
Functionality Compares measurement values with setting values.
How to judge: Based on the display results of the following
table and buzzer sound.
The red backlight turns on, along with the buzzer sounds
when comparison results are WARNING or FAIL.
(When the red backlight turns on, the white backlight turns
off.)
5HVLVWDQFH
ORZ
5HVLVWDQFH
PHGLXP
5HVLVWDQFH
KLJK
9ROWDJHKLJK PASS WARNING FAIL
9ROWDJHORZ WARNING WARNING FAIL
Default OFF