61
Setting Threshold Values for the Comparator
When a measurement value is determined as "FAIL"
FAIL
Resistance
failure
limit
Default
value
Default value
Resistance
warning
limit
Voltage
Resistance
PASS
FAIL
WARNING
Voltage
warning limit
Measurement
value
Comparison Table for the Comparator
The result is determined by the display and buzzer as shown in the
following table:
Resistance
(low)
Resistance
(medium)
Resistance
(high)
Voltage (high) PASS WARNING FAIL
Voltage warning limit
Voltage (low) WARNING WARNING FAIL
Resistance
warning limit
Resistance
failure limit
The boundary conditions are as follows:
Resistance
PASS
Resistance
warning limit
<
Resistance
WARNING
Resistance
failure limit
<
Resistance
FAIL
Voltage
WARNING
<
Voltage
warning limit
Voltage
PASS