Adjusting Zero Value (Zero Adjustment)
Zero Adjustment Troubleshooting
Check items Solution
Is the fuse burned out? If so, replace it with a new fuse. (p. 109)
Do the correction
values obtained exceed
300 counts in either
resistance or voltage
range?
Ensure that the test lead is properly connected
to the instrument.
The test lead may be disconnected. If so,
replace it with a new one.
7U\FOHDQLQJWKH]HURDGMXVWPHQWERDUG
Did you short the test
leads properly while the
instrument is in standby
for correction values?
While the instrument is in standby for correction
values (for approximately 10 seconds), short
WKHWHVWOHDGVXVLQJWKH]HURDGMXVWPHQWERDUG
WRSHUIRUP]HURDGMXVWPHQW
Canceling the Zero Adjustment Operation
Pressing the 0ADJNH\IRUDWOHDVWVHFRQGVZKLOHWKH]HUR
DGMXVWPHQWIXQFWLRQLVDFWLYHFDQFHOVWKH]HURDGMXVWPHQWRSHUDWLRQ
Zero adjustment: ON Zero adjustment: OFF
Press for
at least 2
seconds.