38
Pre-operation Inspection
3.1 Pre-operation Inspection
Before using the instrument, verify that it operates normally to
ensure that no damage occurred during storage or shipping. If
\RX¿QGDQ\GDPDJHFRQWDFW\RXUDXWKRUL]HG+LRNLGLVWULEXWRURU
reseller.
Inspection item Method of checking
Is the fuse burned
out?
7RXFKWKHWHVWOHDGWRWKH]HURDGMXVWPHQWERDUG
If the resistance display still shows [íííí], the
fuse might be burned out or the test lead may be
disconnected. If so, replace it with a new one.
Is the test lead
disconnected?
Is the battery level
VXI¿FLHQW"
At the upper right-hand area of the screen,
indicates the present alkaline battery status. If
is displayed, the alkaline batteries need to
be replaced soon. Be sure to have spare alkaline
batteries available.
Inspecting
batteries
Measurement may not be possible for insulated
WHUPLQDOVGXHWRLQVXI¿FLHQWÀRZRIFXUUHQWIRU
measurement. In such a case, clean the terminal
(remove the insulation) before measurement.