106
Troubleshooting
Error Cause Solution
The comparison
result of the
comparator is not
correct.
The comparator
threshold values are not
set correctly.
Set the correct threshold
values. (p. 57)
Temperature is
not displayed on
the screen when
the model 9460
Clip Type Lead
with Temperature
Sensor is used.
The model 9460
Clip Type Lead with
Temperature Sensor is
not connected correctly.
Connect it correctly.
(p. 53)
Incorrect
measurement
values are
obtained.
or
[
íííí
] is
displayed on the
screen.
or
Both [OVER] and
the maximum
display values
ÀDVKHV
The test lead is not
connected correctly
.
Connect the test lead
correctly. (p. 33)
The test lead is
disconnected.
Replace with a new test
lead.
The fuse is burned out. Replace with a new fuse.
(p. 109)
Zero adjustment is not
performed correctly.
Perform zero adjustment
correctly. (p. 40)
An appropriate range is
not selected.
Select the appropriate
range using the range key.
(p. 39)
The date and time
are completely
wrong when the
instrument is
switched on.
The built-in backup
lithium battery of the
instrument needs
replacement.
The user cannot replace
the battery. Contact your
authorized Hioki distributor
or reseller.