10
Usage Notes
DANGER
Do not short‑circuit two wires to be measured by bringing the clip or jaw
tip of the connecting cords into contact with them. Arcs or such grave
accidents are likely to occur.
To prevent an electric shock, be careful to avoid shorting live lines with the
connecting cords tip.
To prevent a short‑circuit or an electric shock, do not touch the metal part of
the connecting cords tip.
The maximum measurement current varies with the frequency, and the
current that can be measured continuously is limited. Operating the
instrument at less than this limitation is referred to as derating. Do not
measure currents in excess of the derating curve. Doing so may result in
instrument damage or malfunction, a re, or a burn due to sensor heating.
It is recommended to make measurements on the secondary side of the
distribution panel. Making measurements on the primary side of the panel,
where currents are higher, poses a higher risk of instrument or equipment
damage in the event of a short‑circuit.