65
Troubleshooting
Symptom Verication and/or Solution
The current value is larger than
expected.
A current value is displayed even
though there is no input.
The instrument cannot perform measurement accurately in the
presence of a strong magnetic eld from a source such as a nearby
transformer or high-current circuit or in the presence of a strong
electric eld from a source such as a wireless device.
A sound is being emitted by the
instrument’s jaw.
The jaw may emit sound when measuring AC currents in excess of
approx. 500 A, however, there is no effect on the measurement.
The measured value does not
appear.
No measured value is displayed,
even when the connection cords
are shorted.
Insert the connection cords all the way.
Use the proper measurement method.
If no measured value is displayed after attempting the above two
solutions, the instrument may be broken. Have the instrument
repaired.