6
TROUBLESHOOTING
6.10
Imager Duo Operator’s Manual
Note
: If Sakura Tissue-Tek SCA coverslipping film is used, slides must be
cleared with xylene.
Slide Label Format and Placement
The slide label must have the correct accession ID format for the Imager to suc-
cessfully scan and read the ID. Refer to “SLIDE LABELING” on page 3.7.
The slide label must be positioned correctly on the slide so that the ID reader
can locate it.
The slide label must be clean, undamaged and not overhanging the edge of the
slide.