EasyManua.ls Logo

Hologic ThinPrep Imager Duo - SECTION B: Replacing the User-Accessible Fuses

Hologic ThinPrep Imager Duo
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IMAGER DUO MAINTENANCE
5
Imager Duo Operator’s Manual
5.3
CAUTION
The verification slide
is very delicate. Only
handle it by the
edges. Keep it clean
and dust-free at all
times.
Remove the verification slide from the cassette and gently clean it with a lint-
free tissue and alcohol. Inspect the slide to make sure it is not cracked, scratched
or damaged in any way. Contact Hologic Technical Support if the slide is dam-
aged in any way.
Replace the verification slide into slot 1 of the cassette and install the cassette
onto the loading platform of the Image Processor.
Note:
The slide must be placed in slot 1 (the bottom slot) of the slide cassette.
Click the
Store
button on the display screen. Wait while the slide is returned to
the verification slide station. (Figure 5-4.)
Figure 5-4 Please wait While Verification Slide is Stored
The system will return to the idle mode screen when complete.
REPLACING THE USER-ACCESSIBLE FUSES
There are two user-accessible fuses located in the Imaging Station (refer to Fig-
ure 1-5). If the fuses must be replaced, follow these steps:
SECTION
B

Table of Contents

Related product manuals