TROUBLESHOOTING
6
Imager Duo Operator’s Manual
6.11
IMAGING STATION AND IPC ERRORS
Table 6.2: Imaging Station and IPC Errors
Error # Error Text Corrective Action
2407 Slide has been rejected.
See Section C, Slide Preparation
and Quality.
If more than one Imaging Station
is in the lab, insure that the veri-
fication slides are not switched
among Imaging Stations.
2409 Three consecutive 2407 errors
within a cassette generate a
2409.
3106 Slide not found.
Ensure that the slide was prop-
erly loaded in the cassette – with
label facing outward and only
one slide per slot.
If slide was loaded correctly, see
Section C, Slide Preparation and
Quality.
3107 Three consecutive 3106 errors
within a cassette generate a
3107.
3111 Slot has multiple slides.
3113 Slot below has slide out of posi-
tion or multiple slides.
3159 Map Cassette Timeout. • Clean cassette mapping sensor,
refer to page 5.7.
• Power Cycle the Imaging Sys-
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
3303 OCR Read Failed.
See Section C, Slide Preparation
and Quality.
If label was printed on a T3,
replace ribbon.
3304
3305 Invalid Slide ID.
See Section C, Slide Preparation
and Quality.
Call Tech Support to verify the
Slide ID has the proper CRC.