TROUBLESHOOTING
6
Imager Duo Operator’s Manual
6.13
4619 System Error. Consecutive black
frames detected (multiple slides).
Power cycle the system. If the error
persists, contact Technical Sup-
port.
4756 Slide has been rejected.
See Section C, Slide Preparation
and Quality.
Check whether the cell spot is
thick or uneven.
Check whether the slide is
uneven.
4757
4758 Slide has been rejected.
Insufficient fine focus operation.
See Section C, Slide Preparation
and Quality.
4759 Three 4758 errors within a cas-
sette generates a 4759 error.
5050 Three consecutive 4450, 4451,
4452 errors (in any combination)
within a cassette generates a
5050 error.
See errors 4450 through 4475 for
details.
5051 Three consecutive 4453, 4454,
4455 errors (in any combination)
within a cassette generates a
5051 error.
5073 Door(s) Open.
Follow the instructions on the
System monitor. (This may
require power cycling the Sys-
tem.)
Check that the doors are closed.
At User Interface, select CON-
TINUE in the User Assisted
Recovery dialog. If the problem
is resolved, continue processing.
If the problem is not resolved,
power cycle the Imaging Sys-
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
Table 6.2: Imaging Station and IPC Errors