TROUBLESHOOTING
Imager Duo Operator’s Manual
6.13
4619 System Error. Consecutive black
frames detected (multiple slides).
Power cycle the system. If the error
persists, contact Technical Sup-
port.
4756 Slide has been rejected.
•
See Section C, Slide Preparation
and Quality.
•
Check whether the cell spot is
thick or uneven.
•
Check whether the slide is
uneven.
4757
4758 Slide has been rejected.
Insufficient fine focus operation.
See Section C, Slide Preparation
and Quality.
4759 Three 4758 errors within a cas-
sette generates a 4759 error.
5050 Three consecutive 4450, 4451,
4452 errors (in any combination)
within a cassette generates a
5050 error.
See errors 4450 through 4475 for
details.
5051 Three consecutive 4453, 4454,
4455 errors (in any combination)
within a cassette generates a
5051 error.
5073 Door(s) Open.
•
Follow the instructions on the
System monitor. (This may
require power cycling the Sys-
tem.)
•
Check that the doors are closed.
•
At User Interface, select CON-
TINUE in the User Assisted
Recovery dialog. If the problem
is resolved, continue processing.
•
If the problem is not resolved,
power cycle the Imaging Sys-
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
Table 6.2: Imaging Station and IPC Errors
Error # Error Text Corrective Action