TROUBLESHOOTING
6.14
Imager Duo Operator’s Manual
5096 Fail connection to server.
Check switch connections and
data bus harness connections
Power Cycle the Imaging Sys-
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
6615 Slide has been rejected. This is a ‘biological event’ where
the sample may contain abundant
clumpy inflammation or bacteria.
See Section C, Slide Preparation
and Quality.
Manually review the slide.
6616 Three consecutive 6615 errors
within a cassette generate a 6616
error.
6617 Slide has been rejected. See Section C, Slide Preparation
and Quality.
6618 Three consecutive 6617 errors
within a cassette generate a 6618
error.
6619 Vibration Error.
The imaging station is sensitive
to vibrations. Look for vibrating
equipment (such as a vortexer)
near the imaging station.
Be aware of other things (slam-
ming doors, high traffic area)
that may also cause the vibra-
tions.
6628 Slide has been rejected. This is a ‘biological event’ where
the sample may be too dense.
See Section C, Slide Preparation
and Quality.
Manually review the slide.
6629 Three consecutive 6628 errors
within a cassette generate a 6629
error.
6630 Slide has been rejected. If the error persists, contact Tech-
nical Support.
6631 Three consecutive 6630 errors
within a cassette generate a 6631
error.
Table 6.2: Imaging Station and IPC Errors
Error # Error Text Corrective Action