TROUBLESHOOTING
6
Imager Duo Operator’s Manual
6.15
6716 Failed to get a valid batch ID from
the server.
Power cycle the system. If the error
persists, contact Technical Sup-
port.
6901 through
6936
Verification slide and focus errors.
See Section C, Slide Preparation
and Quality.
Clean verification slide.
If more than one Imaging Station
is in the lab, insure that the veri-
fication slides are not switched
among Imaging Stations.
Power Cycle the Imaging Sys-
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
6962 Accelerometer fail.
The imaging station is sensitive
to vibrations. Look for vibrating
equipment (such as a vortexer)
near the imaging station.
Be aware of other things (slam-
ming doors, high traffic area)
that may also cause the vibra-
tions.
6963 Fiducial Mark failure.
Clean verification slide.
• Power Cycle the Imaging Sys-
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
7001 Light Level Drift Outside of Toler-
ance.
Clean verification slide.
May be temperature related.
Verify that the instrument is not
near a heating or A/C duct.
Power Cycle the Imaging Sys-
tem. NOTE that the IS and IPC
should always be power cycled
together. See page 3.29, Shut-
ting Down the System.
Table 6.2: Imaging Station and IPC Errors