OPERATION OF THE IMAGER DUO
3.18
Imager Duo Operator’s Manual
Figure 3-19 Typical Event Log Display
If a slide caused an event during processing, the operator should check it at the
conclusion of the batch to determine if there is any corrective action that would
allow it to be successfully processed in another batch. This may include:
Slide incorrectly loaded into the slide cassette
Slide was not completely dry when loaded into the cassette
Mounting media was on the frosted portion of the slide
Slide contains bubbles
Slide coverslip protrudes over the edge & causes interference
Slide not labeled for use with the Imager Duo