Impact / Impact Pro / Impact Pro IR Operating Instructions
46
The instrument will recognize the Enforcer and display an instruction to 5.
press the 4 button to continue.
Press to continue
Enforcer Calibration
When 3 is pressed, the user should hear the distinctive ‘click’ of the Enforcer gas
delivery solenoid valve.
After about 90 seconds the instrument will display either a pass or fail.6.
or
OK
Press to continue
Enforcer Calibration
Fail
Presstocontinue
Enforcer Calibration
If the instrument reports a failure then check that the gas cylinder has 7.
sufficient pressure. Then repeat once more. Note that the instrument
drawer must be opened and then closed to reactivate the Enforcer. If a
subsequent failure should occur please have the instrument serviced
and calibrated, as it may be due to blocked or dirty filters, or the
sensor(s) has failed.
When completed and successful remove the instrument by pressing 8.
downwards on the rear locking catch and lifting the instrument out. The
instrument will activate all its gas alarms as it detects the gas from the
cylinder, thereby also performing a bump test.
Rear Locking catch
Button
The cylinder should be removed after use.
The instrument automatically switches to a flammable gas setting of methane while
in use with the Enforcer. On completion it will revert to the flammable gas that the
instrument is set to measure. No intervention by the user is required.
5� Fault Finding and Procedures
There are two levels of fault within the instrument.
The first level is a warning, from which the user may be able to recover, e.g. a low
battery. The user must press the 3 button to confirm the warning has been seen.
The second level is usually non-recoverable by the user, and in some cases the
only action available is to switch the instrument off.
In both cases a warning/fault code number is given with a brief description of the
fault, and a contact name/telephone number to enable the user to further assistance
if required. Details of the warning and fault codes are given in Appendix A.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell Impact Series and is the answer not in the manual?

Honeywell Impact Series Specifications

General IconGeneral
BrandHoneywell
ModelImpact Series
CategoryBattery Charger
LanguageEnglish

Related product manuals