Impact / Impact Pro / Impact Pro IR Operating Instructions
6
WARNING
Before each day’s usage sensitivity must be tested on a known concentration of the
gas to be detected equivalent to 25 – 50% of full scale concentration. Accuracy
must be within 0 to +20% of actual. Accuracy may be corrected by calibration.
The Catalytic Flammable Sensors sensitivity can be adversely affected by exposure
to certain substances (silicon and sulphur compounds are examples). Every effort
should be made to avoid exposure to these substances. Following an H
2
S alarm or
repeated gassing with H
2
S a check should be performed on the Catalytic Flammable
sensor to verify its accuracy and a calibration performed if necessary.
By default, no alarms are provided for the 0-100%/Volume Flammable IR sensor.
The user can set up alarm levels via the ICU software.
Hydrogen (H
2
) cannot be detected using the IR principle.
If -0.0% v/v for the electrochemical CO
2
sensor channel is permanently displayed,
a sensor zero calibration needs to be performed in clean air. The alarm level A1
for the electrochemical CO
2
range must not exceed 0.5% v/v.
Do not calibrate the electrochemical CO
2
channel if CO
2
is indicated in clean
air after the apparatus has been exposed to CO
2
concentrations above the upper
limit of the measuring range. In this case leave instrument in clean air overnight
before calibration.
The calibration interval for the electrochemical CO
2
sensor shall not exceed 1
month.
Dispose of the spent cartridge and its packaging in accordance with local
regulations.Donotdisposeofinre.
For WEEE/RoHS information, see the Honeywell Analytics website:
www.honeywellanalytics.com.
IMPORTANT NOTICE
Honeywell Analytics can take no responsibility for installation and/or use of its
equipment if this is not done in accordance with the appropriate issue and/or
amendment of the relevant manual.
The user of this manual should ensure that it is appropriate in all detail to the exact
equipment to be installed and/or operated. If in doubt, the user should contact
Honeywell Analytics for advice.
If further details are required which do not appear in this manual, contact Honeywell
Analytics or their agent.
Note
The failure to observe and abide by the above Warnings and Cautions
may render void the intrinsic safety approval of the Impact Series, and
may remove any right of claim against Honeywell Analytics relating to
product liability or consequential damage to any third party.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell Impact Series and is the answer not in the manual?

Honeywell Impact Series Specifications

General IconGeneral
BrandHoneywell
ModelImpact Series
CategoryBattery Charger
LanguageEnglish

Related product manuals