Item
No. Item Name Description
————————————————————————————————————————
078 MAG LIST 2 ITM 4 Same as Item 76 above.
Default = 014 (Atmospheric Pressure)
079 MAG LIST 2 ITM 5
Same as Item 76 above.
Default = 025 (Press. Transducer Range)
080 MAG LIST 2 ITM 6
Same as Item 76 above.
Default = 044 (Pressure Corr. Factor)
081 MAG LIST 2 ITM 7 Same as Item 76 above.
Default = 043 (Total Corr. Factor)
082 MAG LIST 2 ITM 8 Same as Item 76 above.
Default = 048 (Battery Voltage)
083 MAG LIST 2 ITM 9 Same as Item 76 above.
Default = 031 (Case Temperature)
084 MAG LIST 2 ITM 10 Same as Item 76 above.
Default = 203 (Time)
085 MAG LIST 2 ITM 11 Same as Item 76 above.
Default = 204 (Date)
086 MAG LIST 2 ITM 12 Same as Item 76 above.
Default = 255 (End-of-List)
087 PCOR PRESS UNITS
Code (0-9) that selects the unit of measure for Gas Pressure (008)
and other pressure related items.
Select: 0 - PSIG--Default 5 - mBAR
1 - PSIA 6 - KGcm2
2 - kPa 7 - in WC
3 - mPa 8 - in HG
4 - BAR 9 - mmHG
088 PCOR PRESS DISPL RES
Code (0-4) that selects the number of decimal positions to the right
of decimal point for item 008 and other pressure related items.
Select: 0 - x x x x x x x x
1 - x x x x x x x.x
2 - x x x x x x.x x------Default
3 - x x x x x.x x x
4 - x x x x.x x x x