RAEGuard 3 User Manual
30
Gas type
Flow rate (L/min)
Clean air or high pure nitrogen
0.5 to 1.0
O
2
0.5 to 1.0
H
2
S
0.5 to 1.0
CO
0.5 to 1.0
Flammable gas
0.5 to 1.0
Methane (NDIR)
0.5 to 1.0
The RAEGuard 3 detector adopts a 2-point calibration method, i.e. zeroing
calibration and span calibration. The calibration of the detector should follow the
sequence below.
Note: The zero of the oxygen sensor does not need to be calibrated, nor does it
require bump test. Clean air in the environment (20.9% vol oxygen) can be used
to directly do span calibration.
Zero Calibration
1. Remove the Waterproof Cap and fit the RAEGuard 3 Calibration Cap.
2. Put the magnet over the switch “√” for at least 3s.
3. Click the left switch” ▲” to increment or decrement the values, use the right
switch” ▼” to move to the next character. The basic password is: 1111, the
advanced password is: 8888, select the switch” √” for confirmation.
4. Select the switch” √” to enter the "Calibration & Test" menu as shown below.