EasyManua.ls Logo

Honeywell RAEGuard 3 - Page 31

Honeywell RAEGuard 3
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RAEGuard 3 User Manual
31
5. According to navigation, then select the switch” √” access to the calibration
screen as shown below. At the same time, the detector enters the "Inhibit"
mode.
6. When zeroing gas is ready, please click the switch” √, if it is oxygen sensor,
and “No Zero Calibration” is prompted, then go to “calibration concentration
setting menu. If you give up, please click the switch to return the
“Calibration & Testscreen; click “▼” to skip the zero calibration to enter the
“Span Concentrationsetting menu.
7. Use the switch to abort the zeroing process at any time. After 60 seconds,
if zero is successfully calibrated, click the switch √” to save the results, as
shown below. If not receiving, select the switch” ▲” to abort and return the
“Calibration & Testscreen.
8. When zero calibration is failed, the following screen is as shown below. Click
the switch ”√” to recalibrate zero and repeat above process; click the switch
“▼”, give up and enter the “Span Concentration setting menu; click the
button “▲” to return to the Calibration & Testscreen.

Related product manuals