Page 152 SMV800 Modbus User's Guide Revision 1.0
Simulated Flow value when
Simulation setting is ON
Pay for play option. This
parameter gives information
that what are options present
in transmitter.
Bit 0 = Reserved
Bit 1:-> 0 = Universal
Input Option not present ;1
= Universal Input option
present
Bit 2:-> 0 = Flow Output
Option not present ;1 =
Flow Output Option
present
Bit 3 to 15 = Reserved
Reserve For
Device
Specific
Reserve For
Device
Specific
Reserve For
Device
Specific
Selected
calibration set
for DP
The currently selected
custom factory calibration
(A,B, C) for Differential
Pressure
0 = CAL A
1 = CAL B
2 = CAL C
3 = Best Fit
Available
calibration set
for DP
The available custom factory
calibration (A,B, C) for
Diffrential Pressure
0 = CAL A
1 = CAL B
2 = CAL C
3= Best Fit
Selected
calibration set
for SP
The currently selected
custom factory calibration
(A,B, C) for Static Pressure
0 = CAL A
1 = CAL B
2 = CAL C
3 = Best Fit
Available
calibration set
for SP
The available custom factory
calibration (A,B, C) for Static
Pressure
0 = CAL A
1 = CAL B
2 = CAL C
3= Best Fit
Allows selection of one of the
available custom factory
calibrations for Differential
Pressure
0 = CAL A
1 = CAL B
2 = CAL C
3 = Best Fit
Allows selection of one of the
available custom factory
calibrations for Static
Pressure
0 = CAL A
1 = CAL B
2 = CAL C
3 = Best Fit
Bar Code number of Meter
Body
Configuration option for
Standard or Fast Speed of
Response