Page 204 SMV800 Modbus User's Guide Revision 1.0
Information
Common DB
Corrupt
NVM copy of
Common database
block found corrupt.
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 0
122
Vital Config DB
Corrupt
NVM copy of Vital
Configuration
database block found
corrupt.
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit -1
General Config
DB Corrupt
NVM copy of General
Configuration
database block found
corrupt.
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 2
Config Change
DB Corrupt
NVM copy of
Configuration
Change database
block found corrupt.
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 3
Advance
Diagnostics DB
Corrupt
NVM copy of
Advanced
Configuration
database block found
corrupt.
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 4
Display View
Config DB
Corrupt
NVM copy of Display
View/Screens
Configuration
database block found
corrupt.
Power cycle the
device. If the problem
persists reconfigure
Display Screen
parameters. If still
problem is seen
replace
communication
module.
Bit - 5
Display Common
Config DB
Corrupt
NVM copy of Display
Common
Configuration
database block found
corrupt.
Power cycle the
device. If the problem
persists reconfigure
Display Screen
parameters. If still
problem is seen
replace
communication
module.
Bit - 6
Totalizer Config
DB Corrupt
NVM copy of
Totalizer
Configuration
database block found
corrupt.
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 7