Revision 1.0 SMV800 Modbus User's Guide Page 205
Information
Miscellaneous
DB Corrupt
NVM copy of
Miscellaneous
Configuration
database block found
corrupt.
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 0
123
Totalizer Value
DB Corrupt
NVM copy of totalizer
value block is found
corrupt
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit -1
Flow Unit DB
Corrupt
NVM copy of totalizer
value block is found
corrupt
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 2
Flow Parameter
DB Corrupt
NVM copy of
advance flow value
block is found corrupt
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 3
Backup Totalizer
Value DB
Corrupt
NVM copy of totalizer
value back up block
is found corrupt
Power cycle the
device.
Bit - 4
User Variable DB
Corrupt
NVM copy of user
variables block is
found corrupt
Power cycle the
device. If the problem
persists reconfigure
user defined variables
LRV,URV if
required.If still
problem persists
replace transmitter.
Bit - 5