1-11
General Information
Electromagnetic Compatibility
Electromagnetic Compatibility
This product has been designed to meet the protection requirements of the European
Communities Electromagnetic Compatibility (EMC) directives:
EN55011:1991 (Group 1, Class A)
EN50082-1:1992
- IEC 1000-4-2 (1995) ESD
- IEC 1000-4-3 (1995) Radiated Susceptibility
- IEC 1000-4-4 (1995) EFT
In order to preserve the EMC performance of the product, any cable which becomes
worn or damaged must be replaced with the same type and specification.
Electrostatic Discharge
When an air discharge is applied to the HP37717C in accordance with IEC 801-2
(1991), degradation in performance may be observed in the form of occasional bit
errors being counted.
See also “10 Base-T Lan Connection Radiated Emissions” on page 2-15