1-56
General Information
Specifications
PDH M.2110 error analysis: 2 hour
BIS result, 24 hour BIS result, 7 day
BIS result.
PDH M.2120 error analysis): TX 15
minute threshold report, RX 15 minute
threshold report, TX 24 hour threshold
report, RX 24 hour threshold report.
Alarm indication: AIS, signal loss,
pattern sync loss, frame loss, remote
alarm, multiframe loss, errors present.
Error analysis per ITU-T G.826:
Errored seconds, Severely errored
seconds, Unavailable seconds, Path
unavailable seconds, Errored second
ratio, Severely errored second ratio.
In addition Errored blocks and
Background block error ratio if 2 Mb/s
is selected. If framed all above for
NEAR end and FAR end.
Alarm indication: Signal loss, pattern
loss, AIS, frame loss, multiframe loss
remote alarm/FERF, remote multiframe
alarm.
Alarm seconds: Power loss, AIS,
signal loss, pattern loss, 140 Mb/s
frame loss, 34 Mb/s frame loss,8 Mb/s
frame loss, 2 Mb/s frame loss, RAI,
multiframe loss, remote multiframe
alarm.
Alarm Scan: The interface rate and all
lower levels in the hierarchy are
scanned for occurrences of AIS, frame
loss and RAI. Any occurrence of these
alarms is displayed in graphical form.
Round trip delay: Measures up to 2
seconds delay between transmit and
receive.
Resolution: - 10 s
Option 110 Results
Frequency measurement: Frequency
displayed in Hz, 1 Hz resolution. Offset
displayed in ppm and Hz.
Error results: DS1 (counts & ratios):
Bit, B8ZS/AMI code violations, frame
errors, CRC6 errors.
DS3 (counts & ratios): Bit, B3ZS code
violations, frame errors, P-parity,
CP-parity, FEBE.
E1 (counts & ratios): Bit, HDB3/AMI
code violations, frame errors,CRC4,
REBE.
E3 (counts & ratios): Bit, HDB3 code
violations, frame error.
Alarm indication: DS1:LOS, pattern
loss, AIS, OOF, Multiframe Loss, RAI,
EXZ, Idle
DS3:LOS, Pattern Loss, AIS, OOF,
Multiframe Loss, RAI, EXZ, Idle
E1: LOS, Pattern Loss, AIS, LOF, RAI,
RMFAI, CASMFL
E3: LOS, Pattern Loss, AIS, LOF, RAI
FEAC code indication: With C-Bit
parity framing loopback and alarm/
status codes are decoded and displayed.
Displays shows current and last active
FEAC message.
G.826 analysis: Errored blocks (EB),
errored seconds (ES), severely errored
seconds (SES), unavailability seconds
(UAS), error second ratio (ESR),
severely errored second ratio (SESR),
background block error ratio (BBER),
path unavailable seconds (PUAS).
G.821 analysis (not DS1 and DS3):
EC, SES, %SES, ES, %ES, EFS,
%EFS, unavailability, %unavailability,
degraded minutes, (%)degraded

Table of Contents

Other manuals for HP 37717C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 37717C and is the answer not in the manual?

HP 37717C Specifications

General IconGeneral
BrandHP
Model37717C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals