1-59
General Information
Specifications
TU-2 path BIP (V5), TU-2 path REI,
TU-12 path BIP (V5),TU-12 path REI,
TU-11 path BIP (V5), TU-11 path REI,
bit errors (bulk filled, PDH payload).
AU-3 path BIP (B3), AU-3 path REI,
AU-3 path IEC.
DS1/DS3 error results:
Frame error, CRC6 error (DS1 ESF),
P-bit parity (DS3), C-bit Parity (DS3
CBP framing), REI (DS3 CBP
framing). Bit errors (DS1 and DS3).
Error analysis: To ITU-T G.826
(G.821 and M.2100/2110/2120 for
PDH payload).
Alarm indication: LOS, LOF, OOF,
MS AIS, MS RDI, K1K2 change
AU-3 path AIS, AU-3 path RDI, AU-3
LOP, AU-3 pointer adj
AU-4 path AIS, AU-4 path RDI, AU-4
LOP, AU-4 pointer adj
TU-3 path AIS, TU-3 path RDI, TU-3
LOP, TU-3 pointer adj
TU-2 path AIS, TU-2 path RDI, TU-2
LOP, TU-2 pointer adj
TU-12 path AIS, TU-12 path RDI, TU-
12 LOP, TU-12 pointer adj
TU-11 path AIS, TU-11 path RDI, TU-
11 LOP, TU-11 pointer adj
H4 multiframe sync loss, pattern sync
loss, clock loss, power loss and errors
(any type).
DS1/DS3 alarm indication: AIS, frame
loss, RDI.
Alarm seconds: As for alarm
indication, plus NDF, missing NDF and
except clock loss.
AlarmScan plus alarm and BIP scan:
Automatically scans the SDH network
hierarchy for alarms and BIP errors or
alarms only with a graphical display of
the network hierarchy's status including
the indication of unequipped channels.
Alarms: LOP, path AIS, path RDI, H4
LOM†, TU LOP*, TU path AIS*, TU
path RDI.*
For TU-11, TU-12 and TU-2
structures.
* If applicable.
BIP errors:
AU-4 payloads: VC-4 B3.
AU-3 payloads: VC-3 B3.
TU-3 payloads: VC-4 B3 and VC-3 B3.
TU-2/TU-12/TU-11 payloads:
VC-4/VC-3 B3 and V5 BIP-2.
User selectable BIP error threshold:
Off, > 0, > 10
-3
, > 10
-6
.
Protection switch times:
Service disruption test measures error
burst length for measurement of
protection switch times†.
Accuracy: < 50 ms.
Results: Longest burst length, shortest
burst length, last burst length.
Resolution: 1 ms.
† Service disruption test requires PDH/
DSn option UKJ, UKN or 110 to be
fitted.
Pointer results: AU pointer value, AU
NDF seconds, AU missing NDF
seconds,
AU +ve adjustment count/seconds,
AU -ve adjustment count/seconds, TU
pointer value, TU NDF seconds,
TU missing NDF seconds, TU +ve
adjustment count/seconds,
TU -ve adjustment count/seconds,
implied VC-4, VC-3, VC-2, VC-12,
VC-11 offer.

Table of Contents

Other manuals for HP 37717C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 37717C and is the answer not in the manual?

HP 37717C Specifications

General IconGeneral
BrandHP
Model37717C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals