EasyManua.ls Logo

HP 8590 E-Series - Adjusting the Fine Focus

HP 8590 E-Series
371 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step Keys
The step keys allow discrete increases or decreases of the active function value. The step size
depends upon the spectrum analyzer measurement range or on a preset amount. Each press
results in a single step change. For those parameters with fixed values, the next value in a
sequence is selected each time a step key is pressed. Changes are predictable and can be set
for some functions. Out-of-range values or out-of-sequence values will not occur using these
keys.
Fine-Focus Control
The fine-focus control is located on the side of the spectrum analyzer. Use the following
procedure to adjust the fine-focus control:
1. Adjust the front-panel intensity control for a comfortable viewing intensity.
2. Use an adjustment tool or small screwdriver to access the fine-focus adjustment. See
Figure 2-3. Adjust for a focused display.
F I NE FOCUS ADJUSTMENT
Figure 2-3. Adjusting the Fine Focus
Getting Started 2-9

Table of Contents

Related product manuals