EasyManuals Logo

HT PVCHECKs-PRO User Manual

HT PVCHECKs-PRO
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
PVCHECKs-PRO
EN - 44
4. Use the arrow keys (,) to access the setting of
measurement parameters. The screen on the side
appears on the display. Use the arrow keys (, ) to set
the values. The following options are available:
Man. Set the manufacturer’s name of the module
(max 50) found in the internal DB
Name Set the name of the module (max 50) found
in the internal DB. If, when entered in the database,
the modules has been defined as “Bifacial”, the
instrument and the remote unit will automatically read
3 irradiance values.
N. Mod x STR set the number of modules of the
string in range: 1 ÷ 35
N. Str in par. Set the number of strings connected
in parallel in range: 1 ÷ 10
Mod. Temp set the measuring mode of module
temperature by choosing among the options:
AUTO temperature calculated by the instrument
based on the measured Voc (no probe
connected) – recommended option
MEAS temperature measured via probe PT305
connected to remote unit
MAN manual setting of module temperature if
known in the subsequent range
Tol. Voc set percentage tolerance for Voc
measurement in range: 1% ÷ 15% (typical 5%)
Tol. Isc set percentage tolerance for Isc
measurement in range: 1% ÷ 15% (typical 10%)
Iso V. Test set test voltage for insulation
measurement among the options: OFF
(measurement excluded), 250V, 500V, 1000V,
1500VDC
Iso R.Lim set the minimum reference threshold for
insulation measurement among the values:
0.05,0.10,0.23,0.25,0.50,1.00,50M
RPE Lim set the maximum limit for continuity test
among the values: OFF (measurement excluded),
1,2,3,4,5
5. Press the SAVE key to save settings and go back to the
previous screen.
IVCK 15/10 – 18:04
Man. SUNPOWER
Name: 318WTH
N.Mod. x STR :
01
N.Str.in par. :
02
Mod. Temp :
MAN
55 °C
Tol. Voc :
05 %
Tol.Isc :
10 %
Iso V.Test :
1000 V
Iso R.Lim :
1.00 M
RPE lim :
2