EasyManuals Logo

Idexx VetTest User Manual

Idexx VetTest
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
'o
IDEXX VetTest* Chemistry Analyzer Operator’s Manual
Remove the seal and stopper from the vial. Proceed immediately to running the 3.
quality control procedure on the VetTest analyzer.
Note: Any remaining fluid in the UPRO Control vial is valid for up to seven days as
long as it is kept refrigerated and tightly stoppered. It is important to restopper the
vial and return it to the refrigerator immediately after aspirating the initial contents if
you plan to reuse the control within seven days.
Performing the Quality Control Procedure on the VetTest Analyzer
From the VetTest main menu select 1. 1 for New Sample.
Select 2. 6 for Controls.
The current quality control lot numbers will be displayed. Select the number that 3.
corresponds with the number on your control vial (either VetTrol or UPRO Control).
Proceed as if running a patient sample, following the prompts on the VetTest screen.
Compare your results with the corresponding ranges on the printout.
If one chemistry is out of range, follow this chart to complete the QC procedure:
Check expiration date of slide/control.
Discard expired material.
Repeat same test (x) using same
control and, if possible, another test
in the same optics group.
Proceed with
patient analysis
Run original test (x) with new
box of slides
Proceed with patient analysis
Reconstitute a new control
and repeat tests (x) and (y)
Call IDEXX Technical
Support for assistance
Random Test (x)
Out of Range
First test (x) still out of
range, second test type (y)
within range
Both tests (x) and (y) repeat
out of range
Rerun (x) and/or (y)
out of range
Rerun (x) is still
out of range test
Rerun (x) results
within range
Both tests (x) and (y) repeat
within range

Table of Contents