INFICON Operation CU1000 | 6
CU1000-Operating-Instructions-jina54en1-09-(2404) 23 / 46
Control unit
The control unit settings can be reset to factory settings.
Control unit Functions > Data > Parameters > Reset >
Control unit settings
6.3.4 Recording data
The data is saved as a TXT file. Each TXT file contains the following information:
Date created
Software version
Serial number
Start time
Time stamp (measurement indicates offset in seconds in relation to start time)
File name
Time stamp (offset in seconds in relation to start time)
Leak rate (expressed in selected unit)
Pressure p1 (expressed in selected unit)
Device status
Switching on/off
Switching data recording on/off
Off
On
Control unit Functions > Data > Recorder > Settings >
Data recording
Record interval
Time interval between data recordings
100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s
Control unit Functions > Data > Recorder > Settings >
Record interval
Memory location
The data stored in the control unit can be saved to a USB stick. The memory in the
control unit is limited to the recording of a 24-hour measurement. Each time after one
hour has elapsed, the file is closed and recording continues in the next file.
USB flash drive
Control unit
Control unit Functions > Data > Recorder > Settings >
Storage location
Copy data
Copy data from the internal memory of the control unit to a connected USB stick.
Control unit Functions > Data > Recorder > Copy >
Copy files
Delete data
Delete data in the internal memory of the control unit