6 | Operation CU1000 INFICON
32 / 46 CU1000-Operating-Instructions-jina54en1-09-(2404)
6.3.7.1 Updating the software of the control unit
The software is contained in two files with the same file name, but with different file
name extensions (".exe" and ".key").
1 Copy the files into the main directory of a USB stick.
2 Connect the USB flash drive to the USB port on the device.
3 Select: "Functions > Data > Update > Control unit".
ð Do not switch off the device and do not remove the USB flash drive while the
software is being updated!
4 Check the version information.
5 Select the "Start" button to start the update. Do not switch off the device and do
not remove the USB flash drive while the software is being updated!
6 Follow the instructions on the touchscreen and wait until the update is
complete.
6.3.7.2 Checking and updating the software version of the MSB box
The current software is available from the Inficon support.
The functions of the XL Sniffer adapter set are taken into consideration in system
software version 2.11 or higher.
1 Copy the file with the file name extension ".bin" into the root directory of a USB
stick.
2 Connect the USB flash drive to the USB port on the device.
3 Select: "Functions > Data > Update > MSB".
ð The display shows information on the current and the new software version
as well as on the boot loader.
4 Check the version information.
ð Select the "Start" button to start the update.
ð Do not switch off the device and do not remove the USB flash drive while the
software is being updated! Do not switch off the device and do not remove
the USB flash drive while the software is being updated!
5 Follow the instructions on the touchscreen and wait until the update is
complete.
6 If the system displays warning 104 or 106, confirm with "C".
6.3.7.3 Updating the software of the I/O module
The software of the I/O module can be updated from the control unit if the mass
spectrometer module has at least the software version "MS module 1.02".
1 Copy the file with the file name extension ".bin" into the root directory of a USB
stick.