76 tinb29e1-c (2019-10) VGC083A_B.om
Table "INFICON RS232 and RS485 Command Protocol Summary" (concluded)
READ IG EMISSION
CURRENT
Read the actual IG emission
current being measured by
the gauge
#xxRDIGE<CR>
(e.g.: #01RDIGE<CR>)
*xx_y.yyEzpp<CR>
(eg: *01_1.00E-04<CR>)
When IG is off: *01_0.00E+00
READ IG FILAMENT
VOLTAGE
Read the current IG filament
voltage
#xxRDIGV<CR>
(e.g.: #01RDIGV<CR>)
*xx_y.yyEzpp<CR>
(eg: *01_2.30E+00<CR>)
When IG is off: *01_0.00E+00
READ IG FILAMENT
CURRENT
Read the current IG filament
current
#xxRDIGA<CR>
(e.g.: #01RDIGA<CR>)
*xx_y.yyEzpp<CR>
(eg: *01_3.10E+00<CR>)
When IG is off: *01_0.00E+00
READ IG ION
CURRENT
Read the current IG ion
current
#xxRDIGC<CR>
(e.g.: #01RDIGC<CR>)
*xx_y.yyEzpp<CR>
(eg: *01_4.77E-09<CR>)
When IG is off electrometer offset
voltage is indicated:
GET FILAMENT
SELECTION
Read current filament
selection
#xxRF<CR>
*xx_FIL SEL1<CR>
NOTES:
1. <CR> is a carriage return (0x0D). This is the terminator character.
2. The '_' used in certain defined commands represent a space character. Space
characters may be omitted from commands.
3. Hardware handshake controls do not exist on VGC083 (e.g., RTS, CTS, DTR).