2.10 Warnings
This equipment is for use only by qualified medical professionals trained in
the particular technique to be performed.
Only handpieces manufactured or approved by InMode MD Ltd. should be
used with Optimas System.
Connect the power cord to a properly polarized and grounded power
source with the frequency and voltage characteristics that match those
listed on the back of the unit.
Connect the System power cord to a properly grounded receptacle. Do not
use power plug adapters.
Always turn off and unplug the device before cleaning.
The patient and treatment attendants must use protective eyewear during
the use of laser and IPL.
The patient should not come into contact with metal parts which are
earthed or which have an appreciable capacitance to earth. The use of
antistatic sheeting is recommended for this purpose. Treatment bed or
chair should not be electric.
Use the lowest output setting necessary to achieve the desired treatment
effect. The higher RF or laser/IPL energy is applied, the greater the
possibility of unintended thermal damage.
Failure of the equipment could result in an unintended increase of output
power.
The cables of the Handpiece should be positioned in such a way that
contact with the PATIENT or other leads is avoided.
Fire / Explosion Hazard - The following substances will contribute to
increased fire and explosion hazards in the operating room:
– Flammable substances (such as alcohol-based skin prepping agents and
tinctures).
– Naturally occurring flammable gases which may accumulate in body
cavities such as the bowel.
– Oxygen enriched atmospheres.