Section 11: System Maintenance
67 of 87
Section 11: System Maintenance
11.1 Maintenance & Frequency
The following list suggests how often each of the maintenance procedures should be
performed by the operator.
11.2 Before and after each treatment
Wipe the device with a damp soft cloth. The Handpiece elements that are in contact
with the skin should be disinfected with 70% alcohol between patients, as described
in detail for each handpiece in “Handpiece Cleaning Instruction Prior to Use”. For
applications that require gel (DiolazeXL, Vasculaze, Lumecca and Forma), remove the
gel before cleaning.
11.3 Once a Week
Clean the System at least once a week. Turn the System off and wipe all surfaces. Be
careful not to spill any liquids on the System.
Clean the Dust Cover. In dusty environments or desert regions, it should be checked
daily.
11.4 Once a Month
Replace the cooling water at least once a month. Follow the instruction of “Draining
Water” and “Filling Water” below.
11.5 Once a Year
By the operator:
Replace the Deionizer Cartridge and Capsule Filter at least once a year. Follow
the “Replacing the Deionizer Cartridge” and Replacing the Capsule Filter”
instructions below.
Replacement Deionizer Cartridge and Capsule Filter can be purchased by your
local service center.