To install a new Deionizer Cartridge:
1. Connect the water hose into the Deionizer filter until is securely locked.
2. Remove the water bottle from the compartment, pour out the distilled water
from the bottle according to “Draining Water” procedure and wash it with new
fresh water.
3. Fill 1 liter of distilled water into the bottle and return it to its compartment.
Perform “Filling Water” procedure as described above.
11.10 Replacing the Capsule Filter
For safety, and proper System performance, replace the cooling water with fresh
distilled water only.
To remove the Capsule Filter:
1. Drain the water from the System following “Draining Water” procedure
2. Turn the System off.
3. Push the slide latch of the bottle door down and remove the door.
4. Holding the Capsule Filter with one hand, loosen the filter by pressing the
upper black quick connector and pulling the filter out of the water hose (Figure
11-12 left).
5. Holding the Capsule Filter with one hand, loosen the filter by pressing the
lower black quick connector and pulling the filter out of the water hose
(Figure 11-12 right).
Figure 11-12: Upper (Left) and Lower (Right) Filter Quick Connectors