Section 3: System Installation
3.1 Electrical Requirements
▪ The System will require a separate line supply of single phase (100Vac; 15A) or
(115Vac; 15A) or (230Vac; 15A) or (240Vac; 15A) 50-60Hz. Zmax = 0.03Ω.
▪ Power receptacles must be within 15 feet of the System site.
▪ The System should not share a power line with other equipment.
▪ Power receptacle must include protective earth and must be checked before
connecting the System.
For continued protection against fire, replace the fuse only with one of the
same type and rating.
Proper grounding is essential for safe operation.
3.2 Environmental Requirements
▪ Corrosive materials can damage electronic parts; therefore, the System should
operate in a non-corrosive atmosphere.
▪ For optimal operation of the System, maintain room temperature between
20º-27ºC (68º-79ºF) and relative humidity of less than 80%.