Maintenance
Before switching on the instrument make sure that the mains cable, mains plug, socket and input at
the instrument are in a perfect state. If the drive of the rotary chair becomes noisy the driving belts
should be renewed. We recommend renewing the driving belts after five years.
The operator is responsible for the perfect state of the instrument. It is recommended by Interacoustics to
have safety checks carried out at one-year intervals.
Immediate maintenance is required if:
the instrument was subjected to extreme mechanical stress (impact, defective cable due to
inadmissible traction);
fluid gets into the instrument;
cables and/or connectors are defective;
rubber-joints have developed cracks.
For further details please contact your Interacoustics representative.
SYMBOLS ON INTERACOUSTICS COMPONENTS
Technical Specifications
Maximum speed 200 deg/s
Maximum acceleration 100 deg/s²
Maximum patient weight 135 kilos
Reclining backrest manually operated from 0 deg (horizontal) to 90 deg (upright)
Emergency stop emergency stop button disconnects the motor power
Patient alarm button sends an alarm signal to the computer and turns off power to the
motor
Weight 195 kg
Shipping weight 230 kg
Dimensions 90 cm x 70 cm x 160 cm
Shipping dimensions 100 cm x 120 cm x 190 cm
Power supply 110-120VAC 220-240VAC
50-60Hz 50-60Hz
900VAC max 900VAC max
Ground conductor connection
Unit of type B