EasyManua.ls Logo

Interacoustics VN415 - Impedance Test

Interacoustics VN415
347 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Working with ENG: EO425 Page 27
ENG/DLB Instruction for Use - English
Date: 2011-09-27 Page 27
It is important to note that when switching from a monocular (2-channel) to
a binocular (3-channel) electrode montage within the individual test
configurations, it does not apply any electrode montage changes to the
system configuration. If you wish to make a change to the default electrode montage you must
make the change in the system configuration. This can be done by going to ‘Tools’ and then
‘Switch between ENG/VNG’. A dialog box will then appear where you can select mode as either
monocular or binocular.
Figure 1.11 Switch between ENG/VNG Figure1.12 Select source
If you switch between montages, i.e. if you switch from a 2 channel montage to a 3 channel
montage, you will be prompted to recalibrate, as seen in the message below:
Figure 1.13 repeat calibration
Impedance test
Once the electrodes have adjusted to the patient, you need to do an impedance test. To do this
you open a test or a session, and click the calibration icon :
Figure 1.14 Calibration and impedance check

Table of Contents

Related product manuals